PrimePCR™ SYBR® Green Assay: NKRF, Human

PrimePCR Primer Assays for Real-Time PCR oligo primer pair tube for SYBR Green gene expression

Real-time PCR primer assay designed for SYBR® Green gene expression analysis.

Info:   Same primer pair as used in probe assay qHsaCIP0029680

List Price:    $174.00
Your Price:   Log In
 
Add to Custom Plate

Assay Information

Technology:   qPCR
Assay Type:   SYBR® Green
Application:   Gene Expression
Unique Assay ID:   qHsaCID0015059
Assay Design:   Intron-spanning
Chromosome Location:   X:118725146-118726437question
Amplicon Length:   189
Splice Variants Targeted:   ENST00000371527 ENST00000304449 ENST00000542113

Gene Information

This gene encodes a transcriptional repressor that interacts with specific negative regulatory elements to mediate transcriptional repression of certain nuclear factor kappa B responsive genes. The protein localizes predominantly to the nucleolus with a small fraction found in the nucleoplasm and cytoplasm. Alternate splicing results in multiple transcript variants. [provided by RefSeq Mar 2010]

Gene Symbol:   NKRF
Gene Name:   NFKB repressing factor
Aliases:   ITBA4, NRF
RefSeq:   NC_000023.10 NT_011786.16 NG_016800.1
Ensembl:   ENSG00000186416
Entrez:   55922
Chromosome Mapping:   Xq24

The below validation information is for the Primer Pair only   Download Validation Data (.pdf)


Products used to generate validation data:

Real-Time PCR Instrument CFX384 Real-Time PCR Detection System
Reverse Transcription Reagent iScript™ Advanced cDNA Synthesis Kit for RT-qPCR
Real-Time PCR Supermix SsoAdvanced™ SYBR® Green Supermix
Experimental Sample qPCR Human Reference Total RNA

Summary Data:

R2 0.999600
y-intercept 35.440000
Efficiency 98

Amplification Plot
Amplification of cDNA generated from universal RNA.

Amplification of cDNA generated from universal RNA.

Melt Peak
Melt curve analysis of above amplification.

Melt curve analysis of above amplification.

Standard Curve
Standard curve generated using 20 million copies of template diluted 10 fold to 20 copies.

Standard curve generated using 20 million copies of template diluted 10 fold to 20 copies.

Number Description Download
10039761 PrimePCR™ Assays Quick Guide, Ver B Click to download